Non-Invasive Coronary Angiography: 
CT and MR Techniques

 

Matthew J. Budoff MD1, Stephan Achenbach, MD2,

And Andre Duerinckx, M.D.3

1Division of Cardiology, Saint John's Cardiovascular Research Center,

Harbor-UCLA Medical Center Research and Education Institute, Torrance, CA USA.

2Department of Internal Medicine II, University of Erlangen-Nuremberg, Erlangen, Germany

3Division of Radiology, University Of Texas Southwestern Med. Ctr,

Dallas, Texas USA

 

Address all correspondence including requests for reprints to:

Matthew J. Budoff, M.D.

Harbor-UCLA Research and Education Institute

1124 W. Carson Street, RB2

Torrance, CA 90502, USA

Tel (310) 222-4107

Fax (310) 787-0448

Email: Budoff@flash.net


Background:   Given the invasive nature and high costs of invasive coronary angiography, there is an increased interest in non-invasive methods to image the lumen of the coronary arteries.  Non-invasive coronary angiography has made great strides over the last decade to become a clinically useful tool to augment conventional coronary angiography.

 

Objective:  We sought to provide a comprehensive review of the literature relating to both electron beam angiography, magnetic resonance angiography and spiral computed tomography, currently the three most promising non-invasive methods to visualize obstructions in the coronary tree.
 

Results:  Angiography and the general indications and uses are discussed.  Three new methods to non-invasively image the coronary arteries are reviewed.  Electron beam angiography provides an overall sensitivity of 86% and specificity of 91% for the 476 patients reported in these studies.  Multi-row-detector spiral computed tomography data is limited, but overall sensitivity is 68% and specificity of 83%.  Magnetic resonance angiography, using newer techniques, demonstrated a sensitivity of 71% and specificity of 82% in the 278 patients reported in these studies. 

 

Conclusions:   These techniques show great promise in their ability to non-invasively visualize the coronary segments. Given the current utility of these techniques, we can expect a rapid growth in both the knowledge and experience with non-invasive angiography, leading to a much wider clinical use of these modalities in visualizing the coronary lumens to evaluate obstructive coronary disease.

 

Over the past decade, great strides have been made in cardiac imaging.  The ability to visualize the lumen of the coronary artery, has been at the forefront of these advances.  Selective cardiac catheterization, the reference standard for visualization of coronary artery stenoses, is an invasive procedure with a significant cost and small procedure related morbidity and mortality. An alternative, less expensive, and non-invasive test for use as a diagnostic tool before possible intervention could have a major impact on health care practice and cost containment.  Non-invasive coronary angiography, due to rapid coronary motion, limited arterial size and tortuous course, has been very challenging.  This paper sought to review the benefits, risks and costs of invasive angiography to that of the three non-invasive methods with the greatest clinical experience, electron beam tomography (EBT), multi-row detector spiral computed tomography (MSCT) and magnetic resonance imaging (MRI).

 

Angiography

Visualization of the lumen of the coronary artery, to assess for the need for revascularization, is currently performed by injection of iodinated contrast directly into the coronary arteries.1  Since the contrast resolution is limited with flouroscopy (the imaging method used during angiography), direct enhancement of the blood pool in the artery of interest is required.  This requires direct arterial puncture with catheters advanced up the aorta (in a retrograde fashion). Conventional coronary angiography (CCA) further requires selective cannulation of the ostium of the left and right coronary arteries and, if present, each saphenous vein graft or internal mammary artery graft to obtain optimal selective contrast enhancement.  Percutaneous or cutdown techniques, usually from the femoral or brachial artery, are used for insertion of special intravascular catheters. The purpose of coronary angiography is to define coronary anatomy and the degree of luminal obstruction of the coronary arteries.1  Coronary angiography remains the standard for assessment of anatomic coronary disease.  The limitations of this procedure include the risks (arterial puncture, iodinated contrast and radiation), the need for multiple staff members including a nurse, physician and technologists, and the costs incurred during the procedure and the ensuing observational period. Given these potential complications, the physician must make reasoned decisions on its use based on the anticipated clinical benefit versus the risks and costs of the procedure.

 

Risk of Invasive Angiography

The procedure is associated with a small but definable risk. A 1990 survey by the Society for Cardiac Angiography and Interventions indicated that the total risk of all major complications (including mortality) from coronary angiography is approximately 2%.2 The most common risk is associated with arterial puncture and advancement of the catheter retrogradely up the aorta to the ostium of each coronary artery. This is required three or more times during a typical angiographic procedure. The catheter can dislodge aortic plaque, which can embolize, causing myocardial infarction or stroke. Furthermore, the arterial puncture can cause vessel damage, sometimes facilitating surgical repair.  Serious bleeding and dissection of the wall of the artery are also rare but significant complications. Another risk of CCA is injection of iodinated dye (which is potentially renal toxic, and can also be associated with allergic reactions, including anaphalaxis).  Once the catheters are removed from the artery, observation and patient immobility is usually required for several hours prior to patient ambulation.  Newer techniques are now available to shorten this post-procedural observation period. In 1993, 1,800,000 cardiac catheterization procedures3 were performed.  Given the present trend of increased utilization, it is possible by 2010,  annual use in the use will exceed 3,000,000 cardiac catheterizations.4

 

Costs  

The 1992 mean charge for cardiac catheterization for inpatients younger than 65 years without a diagnosis of acute MI was $10,880, varying by state from a low of $6,400 in Maryland to $17,600 in California.5 The physician charge made up 18% of the total, averaging $2,000 and varying from $1,300 in South Carolina to $2,550 in California.6 Currently, non-invasive angiography is being offered for $1000-1750 (including physician interpretation) in different medical centers. Research on the cost-effectiveness of non-invasive coronary angiography is needed before the optimal use of this procedure in a wide range of clinical circumstances can be determined.        

 

Current Uses

Despite these and other limitations, coronary angiography is the only method currently available for defining the details of the entire coronary endoluminal vascular anatomy, and it provides the reference standard against which other tests are compared.   Coronary angiography is principally used in three clinical situations7: first, to determine the presence and extent of obstructive coronary artery disease (CAD) in a setting in which the diagnosis is uncertain and CAD cannot be reasonably excluded by noninvasive testing; second, to assess the feasibility and appropriateness of various forms of therapy, such as revascularization by percutaneous or surgical interventions; and finally, as a research tool for the assessment of treatment results and the progression or regression of coronary atherosclerosis. However, the 1999 guidelines concluded “in the absence of clinical indications, repeat angiography is both costly and potentially dangerous.”4   Although coronary lesions that reduce luminal diameter <50% are considered hemodynamically insignificant, they are not clinically benign. These lesions may progress either acutely or chronically, and patients with nonsignificant obstructions have significantly more cardiovascular events during follow-up than those with truly normal coronary angiograms.8

 

Potential Role of Non-Invasive Angiogram

While CCA remains the standard of reference for determining the severity of stenosis resulting from atherosclerosis in the coronary arteries, unfortunately, at least 20% of the clinically indicated diagnostic coronary angiographic procedures performed each year reveal no evidence of obstructive coronary artery stenosis, and therefore do not lead to further interventional procedures.9 This problem may be two-fold in women.  A recent report of 9,238 angiograms from five community hospitals demonstrated that 40.4% of women referred for angiography were found to have non-significant disease, nearly twice that of men.10  Since most of the risk of conventional angiography revolves around the need for arterial access, including stroke, myocardial infarction, infection and bleeding, non-invasive methods, approached by intravenous administration of contrast (or no contrast at all), eliminate most of the risk associated with the procedure, including post-procedural observation periods, myocardial infarction, stroke, bleeding, arterial infection and vessel dissection.

 

Difficulties of  Non-Invasive Angiography

Fluoroscopy, the method used to visualize the coronary artery lumen during angiography, has high temporal resolution, but low contrast resolution.  Thus, it requires direct enhancement of the coronary artery blood pool to allow visualization of the lumen.  Non-invasive angiography, without direct injection of contrast into the artery, requires increased spatial resolution to overcome the problems of loss of superb contrast enhancement and temporal resolution.   Furthermore, rapid motion of the coronary arteries, complex and tortuous anatomy, as well as the small size of the vessels, all make non-invasive imaging challenging.  All three techniques rely upon three-dimensional (3-D) reconstruction to assist with visualization.  The CT techniques (EBT and MSCT), rely upon superior spatial resolution and a venous contrast injection.  The venous contrast does not provide as robust an enhancement of the coronary arteries, since the contrast is diluted by mixing with the blood pool.  Magnetic resonance imaging may utilize gadolinium, a non-iodinated contrast agent, to increase the identification of the lumen.

 

ELECTRON BEAM TOMOGRAPHY

Electron beam tomography with electrocardiographic (ECG) triggering has been used for detecting and quantifying coronary artery calcifications (CAC) for more than ten years. 11,12 EBT is established as the “gold standard” for CAC detection.13 While CAC scores correlate well with the total atherosclerotic burden 14,15 and strongly predict future cardiac events,16,17 the amount of CAC does not correlate well with the stenosis severity of a given lesion.18  Thus, a high calcium score does not always impart a tight stenosis.  While plaque burden itself is dangerous regardless of stenosis severity, use of calcium scanning for predicting the need for angioplasty or bypass surgery is limited. 

Electron beam tomography appears well suited for imaging of the coronary artery.   It has a unique combination of high temporal resolution (100 milliseconds per slice) and spatial resolutions (0.7x0.7x1.5mm),13 allowing visualization of small lesions.  Furthermore, ECG triggering allows image acquisition during the slow portion of coronary motion (diastole).19, 20  Contrast-enhanced electron beam angiography (EBA) is an emerging technology with the potential for obtaining essentially noninvasive coronary arteriograms (figures 1-3).  Recent studies have reported contrast-enhanced, ECG-triggered, 3-D EBA for detecting and grading coronary steno sis. 21,22,23,24,25,26,27

Coronary artery EBA, a noninvasive diagnostic procedure for demonstrating coronary artery anatomy, was first introduced in 1995.21  Two studies 27,28 confirmed that coronary EBA with 3D techniques could reveal the lumen of long segments of the major coronary arteries with high correlation to conventional selective coronary angiography (r=0.83).28  More recently, researchers demonstrated that this modality could be used to identify significant coronary lumen narrowing (>50% stenosis) with the sensitivity of 74-92%, specificity of 79-100% and accuracy of 81.2-93.4% (see table 1).  Summary data demonstrates an overall sensitivity of 86% and specificity of 91% for the 476 patients reported in these studies.  Also, inter-observer reproducibility was found to be high (0.74-0.90 by Cohen's Kappa statistic).25,26

There is fair uniformity and imaging standards among EBA studies.  The imaging procedure used by the above studies are all performed with similar imaging techniques. The studies were performed with a C-100 or C-150XLP electron beam CT scanner (Imatron, South San Francisco, California). ECG triggering is employed, so that each image is obtained at the same point in early diastole. Iodinated contrast is administered through an antecubital or jugular vein with an injection rate of 4 ml/sec and total volume of 120-160 ml.  Forty to fifty images are obtained over a single breath hold, usually over 25-40 seconds. This entire protocol is performed within 15 minutes.  The test is easy to perform, and interpretation can be made in minutes.

 

EBT Limitations

EBA studies had a technical success rate of 85-100% and 8-25% of coronary arteries were non-assessable (table 1). Impaired image quality, due to multiple image artifacts including coronary artery motion and breathing artifacts, limited its clinical use.  In EBA studies, distal arteries are not as well visualized as the proximal segments. This is most likely due to the greater motion artifacts from cardiac pulsation and the partial volume effect (due to small vessel diameters) in distal epicardial vessels, as compared to the proximal segments.23,25,29  The main determinant of false positive results for diagnosing ³50% coronary luminal stenosis was small vessel size21,29  and the diameter of stenotic segments tends to be underestimated by EBA.23

Several strides have been made over the last year to improve on image quality and interpretation of the EBA.  Recent studies revealed that the ECG triggering at 80% of the R-R interval (late diastole) used in most prior studies might not be optimal for imaging of the coronary segments near the right or left atria, since atrial contraction during end-diastole causes rapid movement of the base of the heart. 25,30 It has been recently demonstrated that there is less coronary motion at 40% of the R-R interval (early diastole or end-systole),31,32 which might significantly improve imaging.  Improvements including more reliable ECG gating, use of different ECG triggers (40% instead of 80% of the R-R interval) as well as greater experience with this technique will continue to improve accuracy.   Hardware improvements in the EBT scanner should also lead to higher accuracy.  An available high-resolution detector system improves the spatial resolution by almost 40%, thus enabling improved imaging performance for future studies.33 Currently, the high negative predictive values (95-98%) have demonstrated the ability of this tool to effectively ‘rule out’ the presence of obstructive CAD, potentially averting negative invasive coronary angiography in these patients.

 

EBA after Revascularization

EBA is a promising tool in the follow-up after coronary interventions, with the potential to replace some invasive catheterization procedures (figures 4-6). The most common clinical application of EBA is to evaluate patients with symptoms post-coronary artery bypass graft (CABG) surgery.34 The utilization of EBA to detect the patency of CABG grafts has been reported as early as 1986.35 Saphenous vein grafts (SVG), which are generally of large caliber and have little cardiac motion, are especially well suited for non-invasive imaging with EBA (figures 5-6).  Flow studies demonstrate graft patency with sensitivities of 93% to 96% and specificities of 86% to 100%. 36,37,38,39 Using 3-D visualization in patients post-CABG demonstrated a sensitivity of 92-100% and specificity of 91-100% for establishing patency of SVG as compared to conventional coronary angiography.34,40,41 The same studies demonstrated sensitivity and specificity for patency of left internal mammary of 80-100% and 82-100% respectively. Graft aneurysms have also been identified and reported using EBA techniques.42,43

Percutaneous transluminal coronary angioplasty (PTCA) procedures are performed worldwide as one of the main methods for coronary artery stenosis treatment.44  However, even with intra-coronary stents, restenosis (closure) of the site of angioplasty is still the greatest risk, which can lead to both myocardial infarction and death.  Chest pain after stenting or angioplasty often requires visualization of the site of angioplasty to assess for early closure.  A non-invasive method to visualize the site of angioplasty could potentially be used for less typical presentations of acute closure (no typical angina or ECG changes suggestive of ischemia). EBA has been shown to permit imaging of the coronary arteries and detecting high-grade restenosis after PTCA (figure 4). Achenbach et al.45  reported 50 cases that had performed PTCA without coronary stent implantation. The sensitivity and specificity of EBA was 94% and 82%, respectively, to detect severe stenosis (³70% stenosis).

            Intracoronary stenting is now increasingly used to decrease the restenosis rate of PTCA46 and to avoid emergent complications, such as acute thrombus, coronary artery dissection, and even emergent CABG.47  The ability to visualize the coronary lumen through the metal of the stent poses severe problems for EBA, MSCT and MRA.  The widespread utilization of metal stents during revascularization procedures provides a major limitation for the clinical application of these techniques.  One potential solution to the problem of visualizing the lumen within the stented region is to evaluate flow distal to the stent to assess for patency. EBA flow studies have been utilized to document flow distal to the respective stents to assess for stent patency and stenosis.48,49,50,51  Pump et al reported a sensitivity of 78% (18 of 23 stenoses detected) and specificity of 98% (189 of 193 stents correctly judged to be free of stenosis) for the detection of significant in-stent restenosis by EBA flow measurements.

 

Current Clinical Uses

EBA is well suited for post-CABG and PTCA evaluation, pediatric applications, assessment of congenital heart disease and coronary anomalies,52 and allows measurement of wall motion, myocardial mass as well as right and left ejection fractions during the EBA study.53,54 Given the high negative predictive values, use in patients with lower probabilities of obstructive disease will allow physicians to exclude obstructive CAD.  As compared with CCA, EBA is minimally invasive, has an extremely short acquisition time and markedly lower radiation dose.55 Patients unable to breath-hold for 25 seconds, those with significant arrhythmias and morbidly obese patients are poor candidates for EBA.  The Food and Drug Administration (FDA) has approved EBA for non-invasive coronary angiography. 

 

SPIRAL CT

The ‘need for speed’ is essential to obtaining virtually motion-free images.  The significantly improved temporal and spatial resolution of MSCT scanners opens up new possibilities for cardiac imaging.  Due to high in-plane resolution and thin-slice collimation (up to 1.0 mm), high quality images can be obtained.  However, as the mechanical detector head must rotate around the patient, no current CT can obtain a full rotation in less than 500 milliseconds, too slow to allow for motion free imaging of the coronary arteries.56,57  The lack of sufficiently short image-acquisition time requires MSCT studies to obtain many more images (up to 400 images of the coronary tree are obtained), only utilizing a fraction of the obtained data for image reconstruction. Retrospective triggering, acquiring partial data on each scan, may allow shorter acquisition times, but require sophisticated postprocessing algorithms to reduce motion artifacts.57 Several studies, published primarily in abstract form, are listed in table 2.

Achenbach et al57 reported the applicability and image quality of MSCT on 25 patients without coronary stenoses.  This demonstrated that coronary arteries could be visualized over long segments.  On average, 78% of the proximal and mid-segments could be visualized free of motion artifacts, and coronary diameters showed close correlation to quantitative coronary angiography (r=0.86). 

A recent study described 44 patients who underwent MSCT angiography and traditional angiography.  Evaluating all significant obstructions seen by conventional angiography, the overall sensitivity of MSCT in this study was 58%, which varied widely from vessel to vessel.58  The left main coronary artery had a sensitivity of 100%, while stenoses in the circumflex had a sensitivity of only 44%.  Given that patients with heart rates >70 beats per minute were excluded, and only proximal segments were analyzed, results in a general population should be anticipated to be lower.

Another study of 64 consecutive patients who underwent MSCT and coronary angiography was reported.59  In this study, 68% of major coronary arteries could be evaluated. Similar to the previous study, 32 of 58 high-grade stenoses were detected (sensitivity 58%).  In evaluable coronary arteries only, the sensitivity of lesions >50% was 85%, and specificity was 76%.  However, all four major coronary arteries could be evaluated in only 30% of patients.  In reference to this technique the authors note, “its clinical use may presently be limited due to degraded image quality in a substantial number of cases, mainly due to rapid coronary motion”.

A recently presented abstract described 83 patients undergoing angiography and MSCT.121  A total of 72 high grade stenoses were present, with 40 identified by MSCT (sensitivity of 56%).  The specificity in this study was 86%.  Of 332 coronary vessels, 235 (71%) could be evaluated.  The authors concluded, “its (MSCT) clinical use may presently be limited due to degraded image quality in a substantial number of cases, mainly due to rapid coronary motion”.

 

EBT Versus MSCT

            The difference in image acquisition time has lead to marked differences between EBT and spiral CT in regards to CAC scores, with inter-test variability between 40-91% reported.60,61,62 A recent comparison study in 70 patients undergoing both EBT and spiral CT concluded, “spiral CT has not yet proved to be a feasible alternative to electron-beam CT for coronary artery calcium quantification." 63  The lower temporal resolution leads to a lower sensitivity and accuracy for diagnosing coronary calcium, and these differences are also be encountered with 3-D coronary angiography (where motion free images are required to visualize the coronary arteries without major image artifacts).

In the only direct comparison between MSCT and EBA, MSCT 69 displayed a worse signal-to-noise ratio(10.4 vs. 14.0, p < 0.001) meaning that the MSCT image had more noise as compared to EBA.  Furthermore, MSCT was affected significantly more severely by motion artifact than EBT (17% of vessel segments vs. 8% of vessel segments, p < 0.01).64  When evaluating coronary vessels without significant motion artifacts, EBA was able to visualize a significantly greater portion of the coronary tree without motion artifacts than MSCT (94!10 vs 77!15%, p<0.01).  Cumulative data is presented in Table 2.

 

MSCT Limitations

Since radiation is continuously applied while only a fraction of the acquired data is utilized, high radiation doses (doses of 6-10 rad/study) limit the clinical applicability of this modality. 55,69,  In females, the effective radiation doses is another 25% higher than in males (WinDose 2.0a, Scanditronix Wellhofer, Bartlett TN), raising the mean dose from 8 Rads in men to 10 Rads per study in women.67  These radiation doses are two to three times higher than can be expected for conventional angiography (3-4 Rads), and 5-10 fold higher than doses obtained during EBA (1-1.7 Rads).4,65 

The need to image in phases of slow coronary artery motion, which progressively shorten with faster heart rates, poses a significant obstacle to this modality.68  MSCT angiography has noted significant limitations due to cardiac motion have been seen with heart rates >60 beats per minute.69 One evaluation demonstrated that stenosis detection dropped from 90% with slow heart rates (<60 bpm) to 52% for intermediate heart rates (61-90) to 0% for fast heart rates (>90 bpm) with MSCT.59  Current research practice is to inject beta-blockers to slow the heart rate prior to imaging, to increase the length of diastole and slow cardiac motion.  However, slowing heart rates with beta blockers may pose increased risk, in the form of transient bradycardias and heartblock, with potential serious adverse events. So far, the clinical applicability and accuracy of MSCT for the detection of coronary artery stenoses have not been sufficiently validated.70 Vigorous research is now underway to improve this methodology (reduce motion artifacts and radiation dose) and more data should be available in the near future.  MSCT has not yet obtained FDA approval for non-invasive coronary angiography.

 

Magnetic resonance angiography (MRA)

MRA  provides excellent soft tissue contrast, has inherent 3-D capabilities, and allows acquisition in any anatomic plane. Furthermore, MRA does not expose the patient to radiation, nor iodinated contrast, making this the safest of the current non-invasive modalities, with the exception of those patients with pacemakers, implantable defibrillators or recent stent placement. MRA of the coronary arteries became possible in 1991 with the development of a new group of fast MR imaging sequences.71,72  The new MRI techniques may also allow quantification of velocity and flow in coronary arteries.  MR imaging has proved successful in producing angiograms of peripheral vascular anatomy and abnormalities.73  Recent advances in fast MR imaging have allowed for compensation of coronary and respiratory motion.74,75,76,77 However, limitations in spatial and temporal resolution makes visualizing coronary artery lumens more difficult (figure 8-9).

            Several generations of coronary MRA techniques have since been described (Table 3). All techniques use ECG-triggering. First generation breathhold techniques, as described in 1991, acquire one 2-D image per breathhold, however, their clinical usefulness was severely limited.78,79,80,81,82,83, The second generation techniques use respiratory gating or triggering and are referred to as “non-breathholding” or “free-breathing” techniques, as they do not require breathholding during image acquisition.84,85 Although initial implementations were somewhat unreliable, dramatic improvements have since been made.  Third generation techniques allow 3D volume acquisitions in a single breathhold, in combination with real-time interactive slice positioning, appears very promising. Real-time slice positioning, and higher-resolution acquisition schemes, such as spiral MRA, can further improve and facilitate the use of these coronary MR angiographic techniques. 86,87,88,89,90,91,9293,94,95,96,97, (figure 7-8).  The newest versions are now being tested which will allow 3D volume acquisitions in one breathhold and /or real-time interactive slice positioning  (figure 9) 98,99,100,101,102,. In addition the use of MR contrast agents appears to further improve some techniques.103,104 

 

2-dimensional magnetic resonance coronary angiography (first and second generation techniques)

The breath-hold 2-dimensional technique represented the first attempt at high-resolution imaging of the coronary arteries.78  A breath-hold lasting 16 heartbeats could produce a single image (table 3). However, multiple breathholds were necessary to cover all parts of the coronary artery tree, and the exact reproduction of the level of inspiration required extremely good collaboration of the patient and caused problems in up to 44% of all subjects. Also, the selection of proper images required constant input of an experienced investigator throughout the imaging procedure, which typically lasted between 45 and 60 minutes.79 Other limitations of the technique included difficulties in distinguishing veins or pericardial structures from coronary arteries, poor visualization of the left main coronary artery and impaired image quality due to ghosting, ringing and blurring.80  While an early study with MR angiography yielded favorable results,79 multiple subsequent studies could not reproduce the promising results obtained in these initial reports74,80,81,82,83 (table 3).

 

3-dimensional respiratory gated Magnetic Resonance Coronary Angiography Techniques (third generation techniques)

Many problems of two-dimensional MR coronary angiography can be overcome with the acquisition of a 3-D data set ("volume data set"). A contiguous data set can be acquired independently of operator input, and it permits sophisticated post-processing techniques including 3-D reconstructions86,87,88 which may facilitate image evaluation.  To compensate for the long acquisition time necessary to collect the large amount of data required for imaging of the complete cardiac volume, a number of free-breathing techniques have been developed89,90 (figures 7-8). A number of clinical studies have compared the results obtained by these 3D respiratory gated MR coronary angiography techniques to conventional invasive coronary angiography.91,92,93,94  Accuracy varies widely from study to study (sensitivity: 38-83%, specificity: 57-95%) even when the exams of subjects with poor image quality are excluded (table 4).  Summary data reveals a sensitivity of 71% and specificity of 82% in the 278 patients reported in these studies.  However, a major problem of retrospective respiratory gating is the fact that parts of the data necessary to reconstruct the images may not be available in the desired end-expiratory position. In order to fill these gaps it becomes necessary to include measurements which were not made during the ideal respiratory phase, and that may lead to decreased image quality. Therefore, attempts have been made to use prospective gating98 or to "re-register" the volume according to the respiratory shift indicated by the diaphragm position.99  However, these approaches to respiratory gated imaging have not been evaluated in large comparative studies.

 

Limitations

            The combination of temporal and spatial resolution currently available with MRA is limited. For MRA, the spatial resolution is inversely proportional to the temporal resolution.  Currently used protocols use temporal resolution of approximately 125 msec, and spatial resolution of 1.2x1.2x2.0 mm (table 5).  In addition, a relatively low contrast-to-noise ratio, movement artifacts and decreased image quality on the 3-D datasets often prevent adequate visualization of the coronary vessels with the current technology.70 Other limitations are due to calcifications, metal artifacts, thickened pericardium and small pericardial fluid collections, which can mimic the signal void of blood flow.41   Finally, patients with claustrophobia, recently implanted stents or other metallic objects, pacemakers and implantable defibrillators cannot undergo this procedure. 

 

Recent Developments

To overcome the difficulties associated even with 3-dimensional navigator-echo-based image acquisition techniques for coronary MRA, a number of new imaging protocols have been developed. Major aim of these approaches was to increase the signal-to-noise ratio of the coronary artery through the injection of contrast agent (figure 9). Initial reports of such techniques have been published and demonstrated promising results100,103,104,102  A second approach is to use new contrast agents which remain in the intravascular space long enough to achieve signal enhancement during the longer acquisition times necessary for non-breathhold imaging sequences.105,106,  Another aspect in the investigation of coronary arteries by MRI is the analysis of coronary artery flow using phase difference imaging which demonstrates sufficient accuracy in comparison to invasive Doppler flow measurements.107,108 Such studies have been used to assess coronary stents which, due to metal artifacts, cannot be directly visualized by coronary MRA.109  MRI, by use of both flow modes and 3D reconstruction, can also visualize CABG patency with high accuracy.  Studies demonstrate a diagnostic sensitivity of 93-98% and specificity of 85-97% for saphenous vein grafts.110,111

 

EBA versus MRA

In a recent study of 105 patients post-PTCA, EBA revealed significantly higher sensitivity (93%), specificity (65%), positive and negative predictive values than MRA (navigator-echo-based respiratory gated sequences, sensitivity 67%, specificity 49%) for the detection of high-grade restenosis after PTCA.112  A second study of 27 patients undergoing EBA, MRA and conventional angiography was recently reported.  The sensitivity of EBA for significant stenosis was significantly better (77% versus 54% for MRA, p<0.05).  Specificity was high for both EBA and MRA (95% versus 91%, p=n.s.).  EBA visualized more segments of the coronary arteries than MRA (82% versus 74%, p<0.0%) and had higher in plane spatial resolution (EBA 0.7X0.7 mm versus 1.9x1.25x2.0 mm for MRA).113

 

Clinical Applications

            Although the role of coronary artery MRA for stenosis detection has not yet been established, coronary MRA has already been very successful in the detection of coronary artery variants, and the imaging of coronary stents and bypass grafts. Current clinical applications are limited to anomalous coronary artery evaluation114,115 and post-CABG evaluation. The clinical role of MR coronary angiography still needs to be established.116,117  Spatial and temporal resolution are constantly improving and, in selected cases under non-clinical conditions, MR even permits imaging of the coronary vessel wall.76  Multiple new techniques (including new coils, contrast agents and acquisition methodology) are currently being developed  and should continue to improve upon the current state.

 

CONCLUSIONS

            EBA, MSCT and MRI coronary angiography are rapidly developing techniques and currently not an alternative to conventional coronary angiography in all cases (Table 6).  EBA is currently used clinically in certain centers.  Selective use might prove both cost effective and provide a safer, less-invasive method for patients. These non-invasive techniques have potential capabilities of assessing perfusion and coronary flow in addition to coronary anatomy, and thus may provide a comprehensive cardiac evaluation.

Successful development of a non-invasive angiogram with consistently high sensitivity and specificity for obstructive CAD would greatly reduce the cost and also the morbidity and mortality currently associated with conventional coronary arteriography. The replacement of some of these invasive procedures by non-invasive means would be very desirable.  Some potential uses include: following the non-diagnostic stress test; for those persons with intermediate likelihood of CAD (where the step to coronary angiography might be premature); for symptomatic persons post-PTCA and possibly post-stent; evaluating graft patency post CABG, and for early detection of obstructive CAD in the high-risk person.  EBA, MSCT and MRI must prove to be accurate and interpretable in most vessels prior to their widespread implementation in clinical practice. Given the current utility of these techniques, we can expect a rapid growth in both the knowledge and experience with non-invasive angiography, leading to a much wider clinical use of these new techniques in visualizing the coronary lumens to evaluate obstructive coronary disease. 

 

Table 1: Comparisons of contrast-enhanced electron beam CT and invasive coronary angiography

 

Author (ref)

Number

Sensitivity

Specificity

UNEVALUABLE

Achenbach26

125

92%

94%

25%

Budoff25

52

78%

91%

11%

Moshage21

20

74%

100%

--

Reddy24

23

88%

79%

8%

Rensing29

37

77%

94%

19%

Schmermund30

28

83%

91%                

12%

Nakanishi118

37

74%

94%

12%

Moshage101

118

90%

82%

24%

Achenbach119

36

92%

91%

20%

SUMMARY

476

86%

91%

18%

 

 

Table 2: Comparisons of contrast-enhanced multirow detector spiral CT and invasive coronary angiography

Author (ref)

Number

Sensitivity

Specificity

UNEVALUABLE

Achenbach59

64

85%

76%

32%

Knez58

44

58%

91%

30%

Hong120

25

80%

76%

 

Giesler121

83

56%

86%

29%

SUMMARY

216

68%

83%

30%

 

 

Table 3: Coronary MR Angiography technique classification, as proposed by Duerinckx71  and Wielopolski75

 

Coronary MR Angiography Techniques

Generation

Principle

Pros and Cons

First

One slice per breathhold

2D; spatial registration problems; available on all commercial scanners

Second

Free-breathing

3D and  high resolution; but long acquisition times (up to 45 minutes)

Third

3D volume in a single breathhold

3D and low spatial resolution; shorter acquisition times

 

 

Table 4: Sensitivity and specificity for the detection of coronary artery stenosis by 2-dimensional breathhold (first generation) magnetic resonance coronary angiography in comparison to conventional invasive coronary angiography.

 

 

 

 

Author

 

Patients

 

Sensitivity

 

Specificity

 

Spatial Resolution (mm)

 

Temporal Resolution (ms)

 

 

 

Manning79

39

90%

92%

1.4x0.9x5.0

80-104

 

 

Pennell81

39

85%

-

1.6x0.8x5.0

126

 

 

Post82

35

40-63%

89-97%

1.8x1.0x4.0

88-113

 

 

Mohiaddin83

16

56%

82%

1.6x0.8x4.0

126

 

 

Duerinckx74

 

21

63%

-

1.0x2.0x5.0

117

 

 

 

Table 5: Detection of coronary artery stenosis: comparison of the sensitivity and specificity of navigator-echo based respiratory gated and breathhold (third generation) 3-dimensional magnetic resonance coronary angiography to conventional invasive coronary angiography.

 

 

 

Author

 

Patients

 

Sensitivity

 

Specificity

 

Spatial Resolution (mm)

 

Temporal Resolution (ms)

 

 

 

 

 

 

 

 

 

 

Post 91

20

38%

95%

1.2x2.3x2.1

260

 

 

Woodard 92

10

70%

-

1.2x2.0x3.0

-

 

 

Kessler 93

73

65%

88%

1.2x1.2x2.0

125

 

 

Müller 94

30

83%

94%

1.2x1.2x2.0

125

 

 

Sandstede 95

30

81%

89%

1.2x1.2x2.0

230

 

 

Van Geuns113

27

54%

91%

1.95x1.25x2.0

-

 

 

Regenfus122

50

94%

57%

1.4x1.25x1.5

294

 

 

Van Geuns102

38

68%

97%

1.9x1.25x1.5

-

 

 

TOTAL

278

71%

82%

-

-

 

 

 

 

Table 6.  Strengths and Weaknesses of Each of the Three non-invasive angiography techniques.  (Bolded terms demonstrate relative strength)

Electron Beam Angiography

Multi-Slice Computed Tomography

Magnetic Resonance Angiography

FDA Approved

Not FDA Approved

Not FDA Approved

Limited availability

Equipment Widely Available

Equipment widely available

Standardized Protocol

Protocols under development

Protocols under development

Requires Iodinated Contrast and Radiation

Requires Iodinated Contrast and Radiation

No Iodinated contrast or radiation required

Multiple Studies Reported

Limited Studies Reported

Multiple Studies Reported

Consistent Results across different laboratories

Limited experience

Varied results in different laboratories


 

 Figure Legends:

1.   Electron beam angiogram of a person with abnormal nuclear scan (reversible anterior defect).  The three-dimensional reconstruction reveals normal epicardial arteries.  This volume rendered image reveals the left anterior descending (L), Circumflex (C) and Right Coronary artery (white arrow).  Multiple such images, from different angles, are examined for full visualization of the coronary tree.

2.     Electron beam angiogram of the distal coronary arteries.  While the distal right coronary can only be visualized 80% of the time by this technique, this image demonstrates the posterolateral marginal branch (PLMB) and posterior descending arteries (PDA).  Image courtesy of Dr. Beh, Kuala Lumpur, Malaysia.

3.     Electron beam angiogram of the circumflex distribution.  This image demonstrates a normal circumflex and large obtuse marginal. Image courtesy of Dr. Beh, Kuala Lumpur, Malaysia.

4.   Electron beam angiogram taken after symptoms of chest discomfort status post percutaneous coronary angioplasty with stenting of the left anterior artery.  The image reveals a restenosis (white arrow) of the LAD. 

5.      Electron beam angiogram of a person 8 years post coronary bypass surgery.  The Left Internal Mammary graft is widely patent, inserting into the left anterior descending artery (white arrowheads).  The distal artery is well seen and patent, with minimal distal disease.  There are two closed saphenous vein grafts (black arrowheads).  There are two patent saphenous vein grafts (white arrows), one to a diagonal and one to an obtuse marginal.  The right coronary artery has a 100% mid-vessel stenosis (black arrow).

6.      Electron beam angiogram of a person 22 years after coronary bypass surgery.  A saphenous vein graft is widely patent (white arrow), inserting into the left anterior descending artery, just after a high grade stenosis in the native coronary (white arrowhead).  The distal artery is well seen and patent, extending around the apex of te heart. Three saphenous vein grafts are closed proximally (black arrowheads).

7.      Navigator-echo-based, non-contrast enhanced magnetic resonance coronary angiography (A) in a patient with a stenosis of the left anterior descending coronary artery (arrows). (B) Corresponding invasive coronary angiogram of that patient

8.      3-dimensional surface reconstruction of normal coronary arteries obtained by navigator-echo based 3D magnetic resonance coronary angiography.

9.     MR coronary angiogram of the right coronary artery obtained in a single breathhold. A stenosis of the right coronary artery can be seen (arrow, A). (B) corresponding invasive coronary angiogram which also demonstrates the stenosis.

 

 

REFERENCES


[1] . Marcus ML, Schelbert HR, Skorton DJ, et al, editors. Cardiac Imaging: A Companion to Braunwald's Heart Disease. Philadelphia, Pa: WB Saunders, 1991.

[2]  Noto TJ Jr, Johnson LW, Krone R, et al. Cardiac catheterization 1990: a report of the Registry of the Society for Cardiac Angiography and Interventions (SCA&I). Cathet Cardiovasc Diagn 1991;24:75-83.

[3] American Heart Association. Heart and Stroke Facts: 1996 Statistical Supplement. Dallas, TX: American Heart Association, 1996;22.

[4]  Scanlon PJ, Faxon DP, Audet AM, Carabello B, Dehmer GJ, Eagle KA, Legako RD, Leon DF, Murray JA, Nissen SE, Pepine CJ, Watson RM, Ritchie JL, Gibbons RJ, Cheitlin MD, Gardner TJ, Garson A Jr, Russell RO Jr, Ryan TJ, Smith SC Jr.  ACC/AHA guidelines for coronary angiography. A report of the American College of Cardiology/American Heart Association Task Force on practice guidelines (Committee on Coronary Angiography). Developed in collaboration with the Society for Cardiac Angiography and Interventions. J Am Coll Cardiol. 1999 May;33(6):1756-824.

[5] 1992 hospital charges for two major surgical procedures for cardiovascular diseases, II: cardiac catheterization. Stat Bull Metrop Insur Co 1994;75:21-8.

[6] Lee JC, Bengtson JR, Lipscomb J, et al. Feasibility and cost-saving potential of outpatient cardiac catheterization. J Am Coll Cardiol 1990;15:378-84.

[7] Hanley PC, Vlietstra RE, Fisher LD, Smith HC. Indications for coronary angiography: changes in laboratory practice over a decade. Mayo Clin Proc 1986;61:248-53.

[8] Emond M, Mock MB, Davis KB, et al. Long-term survival of medically treated patients in the Coronary Artery Surgery Study (CASS) Registry. Circulation 1994;90:2645-57.

[9]  Johnson LW, Lozner EC, Johnson S, et. al. Coronary arteriography 1984-1987:a report of the registry of the Society for Cardiac Angiography and Intervention. Cathet Cardiovasc Diagn 1989;17:5-10.

[10]  Kugelmass AD, Houser F, Simon AW, et al. Diagnostic results: gender continues to make a difference.  J Am Coll Cardiol 2001; 37: 497A.

[11] Tanenbaum SR, Kondos GT, Veselik KE, et al. Detection of calcific deposits in coronary arteries by ultrafast computed tomography and correlation with angiography. Am J Cardiol. 1989;63: 870-872

[12] Agatston AS, Janowitz WR, Hildner FJ, Zusmer NR, Viamonte M, Detrano R. Quantification of coronary artery calcium using ultrafast computed tomography. J Am Coll Cardiol 1990; 15: 827-832

[13] Wexler L, Brundage B, Crouse J, et al. Coronary artery calcification: Pathophysiology, Epidemiology, Imaging Methods, and Clinical Implications. A statement for Health Professionals. From the American Heart Association. Circulation. 1996; 94: 1175-1192

[14] Breen JF, Sheedy PF, Schwartz RS, et al. Coronary artery calcification detected with ultrafast CT as an indication of coronary artery disease. Radiology 1992;185:435-439

[15] Agatston AS, Janowitz WR, Kaplan G, Gasso J, Hildner, F, Viamonte M. Ultrafast computed tomography detected coronary calcium reflects the angiography extent of coronary arterial atherosclerosis. Am J Cardiol 1994;74:1272-1274

[16] Arad Y, Sparado LA, Goodman K, et al. Prediction of Coronary Events with Electron Beam Computed Tomography. J Am Coll Cardiol 2000;36:1253-60.

[17] Raggi P, Callister TQ, Cooil B, He ZX, Russo DJ, Lippolis NJ, Zelinger A, Mahmarian M. Identification of patients at increased risk of first unheralded acute myocardial infarction by electron beam computed tomography. Circulation 2000;101:850-855

[18] Mautner SL, Mautner GC, Froenlich J, et al. Coronary artery calcification assessment with electron beam CT and histomorphometric correlation. Radiology 1994;192:619-623

[19] Thomas PJ, McCollough CH, Ritman EL. An electron-beam CT approach for transvenous coronary arteriography. J of Comput Assit. Tomogr. 1995;19:383-389

[20] Uchino A, Kato A, Kudo S. CT angiography using electron-beam computed tomography (EBCT): A phantom study. Radiat Med. 1997;15:272-276

[21] Moshage WE, Achenbach S, Seese B, Bachmann K, Kirchgeorg M. Coronary artery stenosis: Three-dimensional imaging with electrocardiographically triggered, contrast agent enhanced, Electron-beam CT. Radiology 1995;196:707-714

[22] Nakanishi T, Ito K, Imazu M, Yamakito M. Evaluation of coronary artery stenoses using electron beam CT and multiplanar reformation. J of Comput Assist Tomogr 1997;21(1):121-127.

[23] Achenbach S, Moshage W, Ropers D, Bachmann K. Curved multiplanar reconstructions for the evaluation of contrast-enhanced electron beam CT of the coronary arteries. AJR1998;170:895-899

[24] Reddy GP, Chernoff DM, Adams JR, Higgins CB. Coronary artery stenoses: Assessment with contrast-enhanced electron-beam CT and axial reconstructions. Radiology 1998;208:167-172

[25] Budoff MJ, Oudiz RJ, Zalace CP, et al. Intravenous three-dimensional coronary angiography using contrast enhanced electron beam computed tomography. Am J Cardiol 1999;83:840-845.

[26] Achenbach S, Moshage W, Ropers D, Nossen J, Daniel WG. Value of electron-beam computed tomography for the noninvasive detection of high-grade coronary artery stenoses and occlusions.  N Engl J Med 1998;339:1964-71.

[27] Chernoff DM, Ritchie CJ, Higgins CB, et al. Evaluation of electron beam CT coronary angiography in healthy subjects. AJR 1997; 169:93-99

[28] Achenbach S, Moshage W, Ropers D, Bachmann K. Comparison of vessel diameters in electron beam tomography and quantitative coronary angiography. Int J Card Imaging 1998;14:1-7

[29] Rensing BJ, Bongaerts A, Van Geuns RJ, Van Ooijen P, Oudkerk M, De Feyter PJ. Intravenous coronary angiography by electron beam computed tomography: A clinical evaluation. Circulation 1998; 98:2509-2512

[30] Schmermund A, Rensing B, Sheedy P, Bell MR, Rumberger JA. Intravenous electron-beam computed tomography coronary angiography for segmental analysis of coronary artery stenosis. J Am Coll Cardiol 1998; 31:1547-1554

[31] Mao SS, Lu B, Oudiz RJ, Bakhsheshi H, Liu SCK, Budoff MJ. Coronary artery motion in Electron Beam Tomography. J Comput Assist Tomogr 2000;24:253-258.

[32] Achenbach S, Ropers D, Holle J, Muschiol G, Daniel WG, Moshage W. In-plane  coronary arterial motion velocity: Measurement with electron-beam CT. Radiology 2000; 216:457-463

[33].McCollough CH, Daly TR. Performance Evaluation of Electron Beam CT Scanners: Comparison of Various Hardware and Software Configurations.  Am J Cardiac Imaging. 1996;10(2)Suppl 1:10.

[34] Achenbach S, Moshage W, Ropers D. Noninvasive, T three-dimensional visualization of coronary artery bypass grafts by electron beam tomography. Am J Cardiol 1997;79:856-861

[35]  Bateman TM, Gray RJ, Whiting JS, Matloff JM, Berman DS, Forrester JS. Cine computed tomographic evaluation of aortocoronary bypass graft patency. J Am Coll Cardiol 1986;8:693-698.

[36]   Bateman TM, Gray RJ, Whiting JS, Sethna DH, Berman DS, Matloff JM, Swan HJC, Forrester JS. Prospective evaluation of ultrafast cardiac computed tomography for determination of coronary bypass graft patency. Circulation 1987;75:1018-1024.

[37] Stanford W, Krachmer M, Galvin JR. Ultrafast computed tomography in assessing coronary bypass grafts. Am J Card Imaging 1991; 5: 21-8.

[38]  McKay CR, Brundage BH, Ullyot DJ, et al. Evaluation of early postoperative coronary artery bypass grafts patentcy by contrast-enhanced computed tomography. J Am Coll Cardiol 1983;2:312-317.

[39]  Dai R, Zhang, S, Lu B, Cao C, He S, Bai H, Jing B. Electron beam CT angiography with three-dimensional reconstruction in the evaluation of coronary artery bypass grafts. Acad Radiol 1998;5:863-867.

[40]  Ha JW, Cho SY, Shim WH, et al. Noninvasive evaluation of coronary artery bypass graft patency using three-dimensional angiography obtained with contrast-enhanced electron beam CT. Am J Roentgenol 1999;172:1055-9.

[41]  Lu B, Dai RP, Jing BL, Bai H, He S, Zhuang N, Wu QY Budoff MJ. Evaluation of coronary artery bypass graft patency using three-dimensional reconstruction and flow study on electron beam tomography. J Comput Assist Tomogr 2000;24:663-670.

[42]  Wildhirt SM, Becker C, Reichenspurner H, Reichart B. Three-dimensional reconstruction of a large venous bypass graft aneurysm supplying 2 coronary arteries. Circulation 2000;101:e165.

[43]  Johnson PR, Truitt TD. Saphenous vein coronary artery bypass graft aneurysm demonstrated by electron beam CT. J Comput Assist Tomogr 1994;18:488-491.

[44] Vandormael MG, Deligonul U, Kern MJ. Multilesion coronary angioplasty: Clinical and angiographic follow-up. J Am Cardiol 1987;10: 246-252.

[45] Achenbach S, Moshage W, Bachmann K. Detection of high-grade restenosis after PTCA using contrast-enhanced electron beam CT. Circulation 1997;96:2785-2788

[46]  Serruys PW, Strauss BH, Beatt KJ, Bertrand MZ, Puel J, Rickards AF, Meier B, Goy JJ, Vogt P, Kappenberger L, Sigwart U. Angiographic follow-up after placement of a self-expanding coronary artery stent. N Engl J Med 1991; 324(1):13-17.

[47] Steffenino G, Meier B, Finci L, Velebit V, VonSegesser L, Faidutti B, Rutishauser W. Acute complications of selective coronary angioplasty: Preview of 500 consecutive procedures. Br Heart J 1988; 59:151-158.

[48] Pump H, Moehlenkamp S, Sehnert C, Schimpf SS, Erbel R, Seibel RM, Groenemeyer DH. Electron-beam CT in the noninvasive assessment of coronary stent patency. Acad Radiol 1998; 5(12):858-62

[49] Schmermund A, Haude M, Baumgart D, et al. Noninvasive assessment of coronary Palmaz-Schatz stents by contrast enhanced electron beam computed tomography. European Heart J 1996;17: 1546-1553

[50] Lu B, Dai RP, Bai H, He S, Jing BL, Zhuang N, Gao RL, Chen JL, Budoff MJ. Detection and Analysis of Intracoronary Artery Stent After PTCA Using Contrast-Enhanced Three-Dimensional Electron Beam Tomography. Journal of Invasive Cardiology 2000;12:1-6.

[51] Pump H, Mohlenkamp S, Sehnert CA, Schimpf SS, Schmidt A, Erbel R, Gronemeyer DH, Seibel RM. Coronary arterial stent patency: assessment with electron-beam CT. Radiology 2000 Feb;214(2):447-452.

[52] Ropers D, Moshage W, Daniel WG, Jessl J, Gottwik M, Achenbach S. Visualization of coronary artery anomalies and their anatomic course by contrast-enhanced electron-beam tomography and 3-dimensional reconstruction. Am J Cardiol 2001;87:193-197.

[53] Baik HK, Budoff MJ, Lane KL, Bakhsheshi H, Brundage BH. Accurate Measures of Ejection Fraction Using  Electron Beam Tomography, Radionuclide Angiography, and Catheterization Angiography. Int J Cardiac Imaging 2000;16:391-398.

[54] Budoff MJ, Brundage BH. Newer Imaging Modalities: Electron Beam Computed Tomography Willerson JT, Cohn JN eds. Cardiovascular Medicine, Second Ediction. Philadelphia: W.B. Saunders Company, 2000, Chapter 21.

[55]  Becker C, Schatzl M, Feist H, Bauml A, Schopf UJ, Michalski G, Lechel U, Hengge M, Bruning R, Reiser M. Assessment of the effective dose for routine protocols in conventional CT,  electron beam CT and coronary angiography. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr 1999;170:99-104.

[56] Lu B, Mao SS, Zhuang N, Bakhsheshi H, Yamamoto H, Takasu J, Liu SCK, Budoff MJ.  Coronary artery motion during the cardiac cycle and optimal ECG triggering for coronary artery imaging.  Invest Radiol 2001;36:250-256.

[57]  Achenbach S, Ulzheimer S, B aum U, Kachelriess M, Ropers D, Giesler T, Bautz W, Daniel WG, Kalender WA, Moshage W.  Non-invasive coronary angiography by retrospectively EcG-gated multislice spiral CT.  Circulation 2000;102:2823-2828.

[58]  Knez A, Leber A,  Becker CR, et al. Multirow-detector-helical computed tomography for non-invasive angiography: a new imaging modality for determination of severe coronary artery disease. Circulation 2000:102:II-411 (abstract).

[59]  Achenbach S, Giesler T, Ropers D, Ulzheimer S, Derlien H, Schulte C, Wenkel E, Moshage W, Bautz W, Daniel WG, Kalender WA, Baum U. Detection of coronary artery stenoses by contrast-enhanced, retrospectively electrocardiographically-gated, multislice computed tomography. Circulation 2001;103:2535-8.

[60] Budoff MJ, Mao SS, Zalace CP, Bakhsheshi H, Oudiz RJ. Comparison of Spiral and

Electron Beam Tomography in the Evaluation of Coronary Calcification in Asymptomatic Persons. Int J Card Imaging 2001: in press.

[61]  Becker CR, Knez A, Jakobs TF, Aydemir S, Becker A, Schoepf UJ, Bruening R, Haberl R, Reiser MF. Detection and quantification of coronary artery calcification with electron-beam and conventional CT. Eur Radiol 1999;9:620-624.

[62]  Carr JJ, Crouse JR, Goff DC JR, D’Agostino RB Jr, Peterson NP, Burke GL. Evaluation of subsecond gated helical CT for quantification of coronary artery calcium and comparison with electron beam CT. Am J Roentgenol 2000;174:915-921. 

[63] Goldin JG, Yoon HC, Greaser LE 3rd, Heinze SB, McNitt-Gray MM, Brown MS, Sayre JW, Emerick AM, Aberle DR.Spiral versus Electron-Beam CT for Coronary Artery Calcium Scoring. 

Radiology 2001 Oct;221(1):213-21

[64]  Achenbach S, Baum U, Ulzheimer S, Kacheliess M, Ropers D, Kalender WA. Comparison of image quality in contrast-enhanced coronary artery visualization by electron beam tomography and subsecond multi-slice CT. Radiology 2000;217(P):373-374.

[65] Knollmann FD, Hidajat N, Felix R. CTA of the coronary arteries: comparison of radiation exposure with EBCT and multi-slice detector CT. Radiology 2000;217(P):364

[66]  Schmermund A, Hellemann HP, Gruschka-Helleman B, Rauwolf M, Erbel R, Silber S. Multi Row spiral CT for quantification of coronary calcification in 1,000 asymptomatic subjects. Circulation 2000:102:II-399 (abstract).

[67]  International Commission on Radiological Protection. Recommendation of the ICRP.  ICRP Publication 60. Oxford: Pergamon Press, 1990.

[68] Ritchie CJ, Godwin JD, Crawford CR, Stanford W, Anno H, Kim YM. Minimum scan speeds for suppression of motion artifacts in CT. Radiology 1992; 185: 37-42

[69] Ohnesorge B, Flohr T, Becker C, Knez A, Kopp AF, Fukuda K, Reiser MF. Herzbildgebung mit schneller, retrospektiv EKG- synchronisierter Mehrschichtspiral-CT. Radiologe 2000:40:111-7.

[70] Achenbach S, Ropers D, Regenfus M, Pohle K, Giesler T, Moshage W, Daniel W. Non-invasive coronary angiography by magnetic resonance imaging, electron-beam computed tomography and multislice computed tomography. Am J Cardiol 2001;88:70E-73E.

[71] Duerinckx AJ. Imaging of coronary artery disease--MR. J Thorac Imaging 2001; 16(1): 25-34.

[72] Duerinckx AJ. Coronary MR angiography. Radiol Clin North Am 1999; 37(2): 273-318.

[73]  Anderson CM, Edelman RR, Turski PA. Clinical magnetic resonance angiography. New York, NY: Raven Publishers, 1993.

[74] Duerinckx AJ, Urman MK. Two-dimensional coronary MR angiography: analysis of initial clinical results. Radiology 1994;193: 731-738.

[75] Wielopolski PA, van Geuns RJ, de Feyter PJ, Oudkerk M. Coronary arteries. Eur Radiol 2000;10:12-35.

[76] Fayad ZA, Fuster V, Fallon JT, Jayasundera T, Worthley SG, Helft G, Aguinaldo JG, Badimon JJ, Sharma SK. Noninvasive in vivo human coronary artery lumen and wall imaging using black-blood magnetic resonance imaging. Circulation 2000;102:506-510

[77] Botnar RM, Stuber M, Kissinger KV, Kim WY, Spuentrup E, Manning WJ. Noninvasive coronary vessel wall and plaque imaging with magnetic resonance imaging. Circulation 2000;102:2582-2587.

[78] Edelmann RR, Manning WJ, Burnstein D, Paulin S.  Coronary arteries: breath-hold MR angiography. Radiology 1990:181: 641-643.

[79] Manning WJ, Li W, Edelman RR. A preliminary report comparing magnetic resonance coronary angiography with conventional angiography. New Engl J Med 1993;328: 828-832.

[80] Duerinckx AJ, Atkinson DP, Mintorovitch J, Simonetti OP, Vrman MK.  Two-dimensional coronary MRA: limitations and artifacts. Eur Radiol 1996;6:312-325

[81] Pennel DJ, Bogren HG, Keegan J, Firmin DN, Underwood SR. Assessment of coronary artery stenosis by magnetic resonance imaging. Heart 1996;75: 127-133

[82] Post JC, van Rossum AC, Hofman MBM, de Cock CC, Valk J, Visser CA. Clinical utility of two-dimensional magnetic resonance angiography in detecting coronary artery disease. Eur Heart J 1997;18: 426-433

[83] Mohiaddin RH, Bogren HG, Lazmin F, Keegan J, Gatehouse PD, Barbir M, Firmin DN, Yacoub MH. Magnetic resonance coronary angiography in heart transplant recipients. Coronary Artery Dis 1996;7:591-897

[84] Paschal CB, Haacke EM, Adler LP. Three-dimensional MR imaging of the coronary arteries: preliminary clinical experience. J Magn Reson Imaging 1993;3:491-500

[85] Li D, Paschal CB, Haacke EM, Adler LP. Coronary arteries: three-dimensional MR imaging with fat saturation and magnetization transfer contrast. Radiology 1993;187:401-406

[86] Cline HE, Thedens DR, Irarrazaval P, Meyer CH, Ho BS, Nishimura DG, Ludke S.  3D MR coronary segmentation. MRM 1998;40: 697-702

[87] Hany TF, Schmidt M, Davis CP, Göhde SC, Debatin JF. Diagnostic impact of four postprocessing techniques in evaluating contrast-enhanced three-dimensional MR angiography. AJR 1998;170: 907-912

[88] Post JC, van Rossum AC, Hofman MBM, Valk J, Visser CA.  Protocol for two-dimensional magnetic resonance coronary angiography studied in three-dimensional magnetic resonance data sets. Am Heart J 1995;130: 167-173

[89] Hofman MBM, Paschal CB, Li D, Haacke WM, van Rossum AC, Sprenger M.  MRI of coronary arteries: 2D breathhold versus 3D respiratory-gated acquisition. J Comput Assist Tomogr 1995;19: 56-62

[90] Wang Y, Rossman PJ, Grimm RC, Riederer SJ, Ehman RL.  Navigator-echo-based real-time respiratory gating and triggering for reduction of respiration effects in three-dimensional coronary MR angiography. Radiology 1996;198: 55-60

[91] Post JC, van Rossum AC, Hofman MB, Valk J, Visser CA. Three-dimensional respiratory gated MR angiography of coronary arteries: comparison with conventional coronary angiography. Am J Roentgenol 1996;166: 1399-404

[92] Woodard PK, Li D, Haacke EM, Dhawale PJ, Kaushikkar S, Barzilai B, Braverman JC, Ludbrook PA, Weiss AN, Brown JJ, Mirowitz SA, Pilgram TK, Gutierrez FR (1998) Detection of coronary stenoses on source and projection images using three-dimensional MR angiography with retrospective gating: preliminary experience. AJR 170: 883-888

[93] Kessler W, Achenbach S, Moshage W, Zink D, Kroeker R, Nitz W, Laub G, Bachmann K (1997) Usefulness of respiratory gated magnetic resonance coronary angiography in assessing narrowings >= 50% in diameter in native coronary arteries and in aorto-coronary bypass conduits. Am J Cardiol 80: 989-993

[94] Müller MF, Fleisch M, Kroeker R, Chatterjee T, Meier B, Vock P (1997) Proximal coronary artery stenosis: three-dimensional MRI with fat saturation and navigator echo. J Magn Reson Imaging 7: 644-651

[95] Sandstede JJW, Pabst T, Beer M, Geis N, Kenn W, Neubauer S, Hahn D (1999) Three-dimensional angiography using the navigator technique compared with conventional coronary angiography. AJR 172: 135-139

[96] Ropers D, Achenbach S, Birke S, Regenfus M, Daniel WG, Moshage W. Noninvasive coronary angiography using contrast-enhanced magnetic resonance imaging and electron beam computed tomography: comparison of the contrast-to-noise ratio. J Am Coll Cardiol 2000: 415A.

[97] Stuber M, Stuber M, Botnar RM, Danias PG, Sodickson DK, Kissinger KV, Van Cauteren M, De Becker J, Manning WJ. Double-oblique free-breathing high resolution three-dimensional coronary magnetic resonance angiography. J Am Coll Cardiol 1999;34:524-531

[98] Danias PG, McConnell MV; Khasiwala VC, Chuang ML, Edelman RR, Manning WJ Prospective navigator correction of image position for coronary MR angiography. Radiology 1997; 203: 733-736

[99] Oshinski JN, Hofland L, Dixon WT, Pettigrew RI.  Magnetic resonance coronary angiography using respiratory gated real-time slice following. Int J Card Imaging 1998;14: 191-199

[100] Goldfarb JW, Edelman R. Coronary Arteries: Breath-hold, gadolinium-enhanced, three-dimensional MR angiography. Radiology 1998; 206: 830-834

[101]  Moshage W, Ropers D, Daniel WG, Achenbach S. Noninvasive imaging of coronary arteries with electron beam tomography. Z Kardiol 2000;89: Suppl 1:15-20.

[102] van Geuns RJ, Wielopolski PA, de Bruin HG, Rensing BJ, Hulshoff M, van Ooijen PM, de Feyter PJ, Oudkerk M. MR coronary angiography with breath-hold targeted volumes: preliminary clinical results. Radiology 2000;217:270-277.

[103] Wielopolski PA, van Geuns RJM, de Feyter PJ, Oudkerk M .  Breath-hold coronary MR angiography with volume targeted imaging. Radiology 1998; 209: 209-219

[104] Kessler W, Laub G, Achenbach S, Ropers D, Moshage W, Daniel WG. Coronary Arteries: MR angiography with fast contrast-enhanced three-dimensional breath-hold imaging – inital experience. Radiology 1999; 210: 566-572

[105] Li D, Dolan RP, Walowitch RC, Lauffer R. Three-dimensional MRI of coronary arteries using an intravascular contrast agent. Magn Reson Med 1998; 39: 1014-1018

[106] Hofman MBM, Henson RE, Kovács, Fischer SE, Lauffer RB, Adzamli K, De Becker J, Wickline SA, Lorenz CH. Blood pool agent strongly improves 3D magnetic resonance coronary angiography using an inversion pre-pulse. Magn Reson Med 1999; 41: 360-367

[107] Hundley WG, Lange RA, Clarke GD, Meshack BM, Payne J, Landau C, McColl R, Sayad DE, Willet DL, Willard JE, Hillis LD, Peshock RM.  Assessment of coronary flow and flow reserve in humans with magnetic resonance imaging. Circulation 1996; 35: 521-531

[108] Kessler W, Moshage W, Galland A, Zink D, Achenbach S, Nitz W, Laub G, Bachmann K. Assessment of coronary blood flow in humans using phase difference MR imaging. Int J Card Imaging 1998; 14: 179-186

[109] Nagel E, Hug J, Bünger S, Bornstedt A, Schnackenburg B, Wellnhofer E, Klein C, Thouet T, Schalla S, Fleck E. Coronary flow measurements for evaluation of patients after stent implantation. MAGMA 1998;6: 184-185

[110]  Galjee MA, Van Rossum AC, Doesburg T, et al. Value of magnetic resonance imaging in assessing patency and function of coronary bypass grafts: an angiographically controlled study. Circulation 1996;93:660-666.

[111] Vrachliotis TG, Bis KG, Aliabadi D, et al. Contrast-enhanced breath-hold MR angiography for evaluating patency of coronary artery bypass grafts. AJR 1997;168:1073-80.

[112] Ropers D, Achenbach S, Birke S, Regenfus M, Daniel WG, Moshage W. Detection of high grade restenosis after coronary angioplasty using non-invasive contrast-enhanced magnetic resonance imaging and electron beam computed tomography in 105 patients. J Am Coll Cardiol 2000: 415A.

[113] Van Geuns RJ, Bongaerts AH, De Bruin HG, Rewnsing BJ, DeFiejter PJ, Oudkerk M. Noninvasive coronary imaging: a comparison between respiratory gated magnetic resonance imaging and electron beam computed tomography. Eur Heart J 1999;20:263

[114] McConnell MV, Ganz P, Selwyn AP, Li W, Edelman RR, Manning WJ (1995) Identification of anomalous coronary arteries and their anatomic course by magnetic resonance coronary angiography. Circulation 92: 3158-3162

[115] Post JC, van Rossum AC, Bronzwaer JGF, de Cock CC, Hofman MBM, Valk J, Visser CA. Magnetic resonance angiography of anomalous coronary arteries. A new standard for delineating the proximal course? Circulation 1995;163-3171

[116]  Yucel EK, Anderson CM, Edelman RR, et al.  AHA Scientific Statement. Magnetic Resonance Angiography: update on applications for extracranial arteries. Circulation 1999;100:2284-2301.

[117] Budinger TF, Berson A, McVeigh ER, Pettigrew RI, Pohost GM, Watson JT, Sickline SA. Cardiac MR imaging: Report of a working group sponsored by the National Heart, Lung, and Blood Institute. Radiology 1998;208: 573-576

[118]  Nakanishi T, Ito K, Imazu M, Yamakido M. Evaluation of coronary artery stenoses using electron-beam and multiplanar reformations. J Comput Assist Tomogr 1997;21:121-127.

[119]  Achenbach S,  Ropers D, Regenfus M, Muschiol G, Daniel WG, Moshage W. Contrast enhanced electron beam computed tomography to analyse the coronary arteries in patients after acute myocardial infarction. Heart 2000;84: 489-93.

[120] Hong C, Becker CR, Knez A, Schoepf UO, Bruening RD, Reiser MF.  Multislice CT coronary angiography for segmental analysis of coronary artery disease. Radiology 2000:217;374 (abstract).

[121]  Giesler T, Ropers D, Baum U, Ulzheimer S, Schulte C, Achenbach S.  Contrast-Enhanced, Retrospectively ECG-gated multislice spiral CT: Detection of coronary stenoses.  Radiology 2001;221(abstract).

[122] Regenfus M, Ropers D, Achenbach S, Kessler W, Laub G, daniel WG, Moshage W. Noninvasive detection of coronary artery stenosis using contrast-enhanced three-dimensional breath-hold magnetic resonance coronary angiography. J AM Coll Cardiol 2000;36:44-50.